Error Cause Repair
Out of range of
impedance
When the subject’s body impedance deviates from
the limit
-
Insufficient touch to electrodes
Impedance is out of range
Range: 100 ~ 950 Ω
• Clean the measuring parts (the electrodes, palms, and soles) and try again.
• Measure again with correct posture.
• Do not move during measurement.
• If the same error is repeated, please contact our company or its local distributor from
where this device is purchased.
When the subject’s P.B.F. deviates from the limit
-
Incorrect input of personal data
P.B.F. is out of range
• Clean the electrode holders and try again.
• After checking that there is neither something with wrong input of personal data (age,
gender) nor with measuring error of weight and impedance, try again.
• It can’t measure if the P.B.F. is out of range.
•
When the same error occurs even after re- measurement, please contact our company or
its local distributor from where the device was purchased.
Out of range of
measureme nt
When the subject’s fatness is deviated
from the limit
-
Mechanical error
• Input height correctly or if installed height already, measure again.
• Confirm to measure weight and try again correctly.
• It can’t measure if the fatness is out of range.
• When the same error is occurred even re
- measurement, please contact with our
company or its local distributor where is purchased.
Insufficient
electrode
connection
When even one of all 4 electrodes connects with
ankle insufficiently
• Connect to ankle all 4 electrodes once more.
ERROR & REPAIR