15
exactmeters.com
COPPER
FILL, DIP, READ
5
REMOVE STRIP
Remove one eXact
®
Strip Micro CU (Part No. 486629)
and set in a dry, convenient place. Replace cap.
6
DIP STRIP AND PRESS READ
Press to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
strip in the sample,
gently touching the bottom of the cell. Use a gentle
constant back and forth motion (2 strokes per
second) until the timer displays
“1”.
Be careful to not
spill the sample from the CELL. Remove and discard
the strip. The display will automatically start to count
up for 120 seconds.
7
CAP CELL AND READ RESULTS
Place the Cell Cover onto the CELL and
READ result displayed as Copper. This result
is automatically stored in the CU(7) menu.
After testing is complete, rinse the sample cell
immediately and clean with brush.