CYANURIC ACID
FILL, DIP, READ
Note S/N on back of photometer before beginning test.
1
POWER ON PHOTOMETER
Press the button to power on the Pool eXact
®
EZ.
2
FILL CELL
Before testing, rinse CELL and clean with brush
thoroughly. Finally rinse the cell 3 times with the water
sample to be tested, then
FILL
cell to capacity to
begin test.
3
SELECT TEST
Press and re-press the button until the display
shows CY(6).
4
CAP CELL AND ZERO METER
Place the Cell Cover onto the CELL and press
and the photometer display reads 0
PPM
, indicating the
meter is ready for testing.