BH3 Series (02024013) ZSME300000289
EXCELL PRECISION CO., LTD.
Load the indicated calibration
weight onto the platter
according to the
display. Press
the key
3-2-1 External Weight Calibration
The input calibration weight must be the value ranged from 90% to 110% of
interval value of the original calibration weight.
The display shows the calibration weight
(equal to the full capacity) and the weight
unit. The first digit flashes.
The calibration weight can be
modified if necessary.
The display flashes the calibration
weight to be loaded.
Ensure the platter is on the
balance and make sure there
is no weight on the platter.
Press the
key.
When the scale reaches stability, the
display shows calibration weight.
Unload all the weights from the platter, and
the calibration procedure is completed.
Upward digit selecting (0-9)
Downward digit selecting (9-0)
Move the cursor leftward
Move the cursor rightward
Confirmation key