2-3-2 Counting Mode
Sampling
1. Press the key to select the Counting mode and the display shows pcs.
2. Press the key to select a sample size from 10, 20, 50, 100, or 200 pcs.
3. Place as many items as the selected sample size, and press the key. After the sign
“ “disappears, the sampling process is completed and the balance is ready to
count.
4. Press the / key to shift between “Unit Weight Mode Display” and “PCS Mode
Display”.
The unit weight of sample is not sufficient: The weight of a sample should be heavier than
the 0.2d (d=division). When the unit weight of the sample is too small (less than 0.2
divisions), “ pcs” is displayed.
The last sample size is memorized when the balance is switched off. When the balance is
switched on again, press the
key to enter the counting mode.
If “Auto Unit Weight Averaging” setting is enabled, the scale will recalculate average unit
weight when current sample count (of current objects on the platter) exceeds previous
sample count by more than 5 pcs and is less than 200% of previous sample count, i.e.
(p + 5) < (c) < (200% x p)
whereas
p = previous sample count
c = current sample count