02025100 ZSME300000340
EXCELL PRECISION CO., LTD.
Quick Setup Calibration
This page is to quickly initiate the scale, for the other functions configuration,
you can refer the chapters below.
Instructions:
Step 1
Power off the scale, find the switch SWA1 on the bottom of machine.
Switch SWA1 to the ADJ position and When power is off, hold fTAREf key and
press fON|ZEROf key. Release fON|ZEROf key and fTAREf key in order. The
display shows .
Step 2
Refer to the chapter 3-1 at page 11 to complete Capacity Configuration.
Step 3
Refer to the chapter 3-4 at page 16 to complete SPAN Calibration.
Step 4
Refer to the chapter 3-2 at page 13 to complete Weight Calibration.
Step 5
When done the initiation, switch the switch SWA1 back to the LOCK position.
If the switch SWA1 is switched to the LOCK position during calibration, the machine
will exit the service mode automatically.