Calibration
CO2 Calibration
1. Press
” to turn the meter on.
2. Place the probe in a known CO
2 reference for 10 minutes.
3. Press the “SET” button 5 times until “C-01” is displayed in the second display.
4. Press ” or “” to select digits to adjust (flashing).
5. Press ” or “”to adjust the display to the reference value.
6. Press ” to store the value and exit the calibration mode.
Humidity Calibration
1. Press
” to turn the meter on.
2. Place the probe in a known humidity reference for 60 minutes.
3. Press the “SET” button 6 times until “SET” and “CAL” are displayed on the LCD.
4. Press ” or “” to adjust the display to the reference value.
5. Press ” to store the value and exit the calibration mode.
Temperature Calibration
1. Press
” to turn the meter on.
2. Place the probe in a known temperature reference for 60 minutes.
3. Press the “SET” button 7 times until “SET” and “CAL” are displayed on the LCD.
4. Press ” or “” to adjust the display to the reference value.
5. Press ” to store the value and exit the calibration mode.

Other manuals for Extech Instruments EA80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Extech Instruments EA80 and is the answer not in the manual?

Extech Instruments EA80 Specifications

General IconGeneral
BrandExtech Instruments
ModelEA80
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals