Setupmode
Note:IftheHOLDorMIN/MAXfunctionsareactivetheSetupmodecannotbeaccessed.
ToentertheSetupModepressandholdtheSETTINGbuttonforatleast5seconds(untilthebeep
isheard)andthenusetheSETTINGbuttontoscrollthroughthesetupoptions.Theoptionsare:
SEtAutomatictemperaturecompensationpercentage;settablefrom0.00to5.00%per°C
(thedefault
is2.00%).Setto0.00%totakeuncompensatedConductivitymeasurements.
Whenpoweriscycledthisparameterwillreverttothedefault2.00%
°C/FSelectthetemperatureunits(°C°F)
OFF/ONAutopoweroffdisable/enable
Continuewiththeinstructionsbelow:
TemperatureCompensationPercentSetting(SEt)
Whenthedisplay
showsSEtpressENTER.SEtwillflashand theupperdisplaywillshowthe
temperaturecompensationpercent.Usethe▲ and ▼ buttonstosetthedesired
temperaturecompensationpercentage.Afterselectingthedesiredvalue,pressENTERto
savethedata.Thedisplaynowshowsthetemperatureunits.Continuebelow:
Temperature
UnitsSelect
ThelowerdisplayshowsCorF.Usethe▲buttontoselectthedesiredunitofmeasureand
thenpressENTERtosave.ThelowerdisplayshouldnowshowOFFYESorNO.Continue
below:
AutoPowerOFFEnable/Disable
ThedisplaywillshowOFFYESor
OFFNO.Usethe▲buttontoselectYESorNO.
YES:AutopowerOFFisenabled(meterswitchesoffautomaticallyafter10minutes)
NO:AutopowerOFFisdisabled
PressENTERtosavetheselection.Themeterwillreturntothenormaloperatingmode.