EasyManua.ls Logo

Extech Instruments MF100 User Manual

Extech Instruments MF100
10 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image

USERGUIDE
AC/DCMagneticMeter
ModelMF100

Question and Answer IconNeed help?

Do you have a question about the Extech Instruments MF100 and is the answer not in the manual?

Extech Instruments MF100 Specifications

General IconGeneral
Data HoldYes
Max/MinYes
Average/CurrentYes
Power Supply9V battery
TypeThermometer
Air Velocity Range0.4 to 30 m/s
Air Temperature Range0°C to 50°C (32°F to 122°F)
Resolution0.1%
Resolution (Air Velocity)0.1 m/s
Accuracy (Moisture)±3%RH
Accuracy (Air Velocity)±3%rdg + 0.2m/s
Unitsm/s, °C/°F
Resolution (Temperature)0.1°C / 0.1°F

Summary

Introduction

Features

Lists key capabilities, including DC/AC measurements, pole indication, and temperature compensation.

Safety Information

Safety Symbols and Notes

Explains safety symbols, battery handling advice, and critical usage cautions for the meter.

Device Components

Meter Description

Identifies and labels the main parts and controls of the Extech MF100 meter.

Probe Description

Identifies and labels the sensing head, handle, cable, and cover of the meter's probe.

User Interface

Display Indicators

Explains the meaning of indicators like N/S, units (mT/G), AC/DC status, and the magnetic reading.

Meter Buttons

Details the function of each button including POWER, HOLD, REC, ZERO, and MODE/DC/AC/SET.

Operation

Performing Measurements

Outlines the sequence for turning on the meter and selecting measurement units and modes.

Measurement Details

Field Indication and Units

Explains how the display shows DC (N/S) and AC fields, along with measurement units (mT/G).

Data Hold and Recording

Describes how to use the Data Hold function and the REC button for MAX/MIN readings.

Settings and Power Options

User Settings Configuration

Explains how to manage auto power-off and select magnetic field units (mT or G) via the SET button.

AC Power Adapter

Details the use of an optional 9V AC power adapter, which permanently powers the meter.

Maintenance and Reset

Factory Reset Procedure

Instructions for resetting the meter to factory default conditions if it becomes inoperative.

Battery Replacement Guide

Step-by-step guide on how to replace the 9V battery when the low battery icon appears.

Technical Specifications

Related product manuals