Error Troubleshooting
1
The jar connected to the FagronLab™ BASIC is not a
FagronLab™ jar.
The jar has to be substituted for a FagronLab™ jar.
2
The FagronLab™ jar lid was screwed alsant onto the jar
and the stirrer cants on insertion.
Open the jar and screw the lid on correctly.
3
At first time extraction the ointment leaks out in a
gush.
Diminish air before the first time extraction.
4 Liquid leaks out on the bottom of the jar.
The bottom of the jar was not completely covered with
foundation all the way to the rim of the jar bottom.
Ensure to cover the sealing lip of the bottom with
foundation.
After using the jar in a water bath, water accumulated
in the cavity on the bottom side of the jar. The water
cannot penetrate into the jar itself through the sealing
lip, remove the excess water with a paper towel.
5 The jar does not attach into the lift arm.
The jar is not an original FagronLab™ jar. Always use an
original FagronLab™ jar.
Please observe the correct handling of the quick-lock
coupling.