FagronLab™ BASIC – User Manual | 40
Functional testing
Assessment OK Not OK
The device is securely placed
Power is supplied
Turn on the FagronLab™ BASIC device with the power switch
The display shows the device type "FagronLab™ BASIC"
The display shows the number of current number of preparations "mixed
<<---->>"
The display shows "OQ - Mode >> >>"
Attach the test mixing unit (review chapter 4) into the quick-lock coupling
(review chapter 5)
The compounding button starts the OQ-Mode
The display shows all the levels during OQ-Mode
The display shows OQ-Mode "OK"
Remove the test mixing unit (review chapter 5)
Observe the operation manual
Performance of the test
Date Signature
Approval by pharmacist
Date Signature