EasyManua.ls Logo

Fasep V585.2.U - Calibration of Wheel Balancer; How to Calibrate the Wheel Balancer (USER)

Default Icon
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Fasep 2000 srl Rev. 1.6
Videotronic V585.2.U: User’s Manual 14 march 2022
2-5
Fig. 42
Fig. 43
Fig. 44
Fig. 45
4 CALIBRATION OF WHEEL BALANCER
4.1 How to calibrate the wheel balancer (USER)
NOTE: the following symptoms indicate need for calibration::
a) check calibration program fails. b) constant low or high weight readings.
c) indicated point of unbalance constantly wrong d) more than 2 spins required to balance wheels
repeatedly.
Switch on the wheel balancer.
Select SET UP > CALIBRATION > SET/OK >
CALIBRATION > SET/OK.
Spin with no wheel on shaft (Fig. 42)
Close the wheel guard or press <START>.
At the end of the spin, put a wheel (Fig. 43)
and close the wheel guard or press <START>.
At the end of the spin, put the calibration
weight (Fig. 44) and close the wheel guard or
press <START>.
NOTE technical calibration is only available for qualified personnel.

Related product manuals