9 Recertification and recalibration of AFGUARD
®
AFGUARD
®
will be delivered calibrated to the customer's requirements. Standard calibration
range is 0 to 50 pm free water in Jet Fuel. Other calibrations ranges are possible, but must
be especially requested.
After a certain period of use, AFGUARD
®
needs to be checked to assure actual
measurement results are still in the expected measurement range. A so-called recertification
needs to be done. If recertification fails (deviation of measurement result is out of scope)
the AFGUARD
®
must be recalibrated.
The following recertification intervals are to be followed and are mandatory:
AFGUARD
®
for first installation (brand new AFGUARD
®
):
Two years after initial calibration see calibration report
AFGUARD
®
for second and following recertification intervals:
Every year after previous recertification / calibration
9.1 Firmware on AFGUARD
®
AFGUARD® runs a firmware which must be in line with the requirements defined in JIG
statements. The actual firmware to be conformant to JIG 130 is firmware 1.19.
If you are not sure about your actual firmware status please look at the latest calibration
report of your AFGUARD® or contact your FAUDI-Aviation service organisation.
For further information regarding recertification / recalibration, please contact your local
FAUDI Aviation sales contact.