OPERATION ST100A Series Flow Meter
70 Fluid Components International LLC
Command 151: Read Calibration Flow Limits
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Flow Lower Limit Value
4–7
Float
Flow Upper Limit Value
Response Codes
See Table 17, page 74, for response code list.
Note: 1. S
end Command 159 with no data to reload the instrument’s factory default programming.
Command 154: Read Calibration Temperature Limits
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Temperature Lower Limit Value
4–7
Float
Temperature Upper Limit Value
Response Codes
See Table 17, page 74, for response code list.
Command 157: Read Calibration Pressure Limits
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–3
Float
Pressure Lower Limit Value
4–7
Float
Pressure Upper Limit Value
Response Codes
See Table 17, page 74, for response code list.
Command 159: Write Factory Restore
Byte
Format
Description
Request Data Bytes
1
None
None (no data)
Response Data Bytes
None
None (returns nothing)
Response Codes
See Table 17, page 74, for response code list.
Command 160: Write (4-20 mA) Output Channel #1 Parameters
Byte
Format
Description
Request Data Bytes
0–1
Unsigned-16
D/A Setting for 4 mA Out (OUTZ1)
2–3
Unsigned-16
D/A Setting for 4 mA Out (OUTF1)
4
Unsigned-8
Channel #1 Out Variable
Response Data Bytes
0–1
Unsigned-16
D/A Setting for 4 mA Out (OUTZ1)
2–3
Unsigned-16
D/A Setting for 4 mA Out (OUTF1)
4
Unsigned-8
Channel #1 Out Variable
Response Codes
See Table 17, page 74, for response code list.
Command 161: Read (4-20 mA) Output Channel #1 Parameters
Byte
Format
Description
Request Data Bytes
None
Response Data Bytes
0–1
Unsigned-16
D/A Setting for 4 mA Out (OUTZ1)
2–3
Unsigned-16
D/A Setting for 4 mA Out (OUTF1)
4
Unsigned-8
Channel #1 Out Variable
Response Codes
See Table 17, page 74, for response code list.