FERNO/MAN/01/18/159/UK 2 Pegasus Trolley
2. Checking the trolley
The following checks are recommended before the start of each shift's use of the
Pegasus Trolley.
1. Does the trolley have a properly installed mattress and patient restraining
system? Check the mattress for damage; if any damage is found, it must be
replaced. (Please refer to page 12 for fitting instructions)
2. Do the trolley cot sides raise, lower and lock in position properly?
3. Check that the wheels and brakes operate satisfactorily.
4. Does the trolley lower and raise satisfactorily?
5. Does the ambulance have a properly installed approved locking device?
6. Check the operation of the push/pull handles and ensure the head end handle
clips in the stowed position.
7. Does the headrest extension operate correctly and lock in position.
8. Study this manual before use. Practice without a patient until totally competent in
using the trolley.
Once these checks have been completed, then the trolley is ready to be used in
regular service.
SAFE WORKING LOAD
300 kg in lowest position
200kg in highest position
(Tested in accordance with the requirements of BS EN 1789: 2007)
WARNING
DO NOT OPERATE THE HEIGHT ADJUSTMENT PEDAL WHEN
THE TROLLEY IS FIXED INTO THE VEHICLE LOCK.
THIS WILL CAUSE THE TROLLEY TO JAM IN THE LOCKS AND
MAY CAUSE DAMAGE