Maintenance and Troubleshooting 4-7
Document 40275-0913
4.5 Verifying That the FT-170/CI Auto
Enable/Disable Monitor Works Correctly
4.5.1 Steps
Only the FT-170/CI includes the auto enable/disable monitor,
which is synchronized to generate the X-ray inspection beam
when the conveyer belt is started and the disable the X-ray beam
when the conveyer belt is stopped.
The red light is on the same electrical circuit as the X-ray
inspection beam generator so you can be certain that if the red
light is on, the X-ray is on and if the red light is off, the X-ray is
off.
Complete these steps to verify that the auto enable/disable
monitor is working correctly.
1. Stop the conveyor.
2. Does the red safety light turn off?
If yes, go to the next step, as the auto enable/disable
monitor is working correctly.
If no, discontinue using the FT-170/CI and immediately
contact your service department, as the auto enable/disable
monitor is not working correctly.
3. Start the conveyor.
4. Does the red safety light turn on?
If yes, stop, as the auto enable/disable monitor is working
correctly.
If no, discontinue using the FT-170/CI and immediately
contact your service department, as the auto enable/disable
monitor is not working correctly.