Supplement to User Manual - ENGLISH Modbus Server
SU_ModbusV2-8, 2015-03-02 65
Tab. 7.8: Offset of the state of the diagnostic values
state of the diagnostic values (integer 16 bit) offset
SNR 0
SCNR 1
VariAmp 2
VariTime 3
detection rate 4
diagnostic error bits 5
Example: measuring channel: B
physical quantities: VariAmp,
VariTime
Modbus register start address: 16400 + 2 = 16402
Modbus register range: 16402...16403
register number: 2