5 Camera Controller
A6000 and A8500 Series User’s Manual
36
The Trigger Time can be set to activate on a very specific time as measured by the real time clock in
the camera.
Trigger Sources
Internal
The trigger is generated internally to run at the frequency set by
the user (Hz).
External
The trigger is generated externally through the Trigger In
connector on the camera rear chassis. (3.3V LVCMOS)
Software
The trigger is generated via a software button (Trigger button)
Time
Camera generates an internal trigger when the internal
timestamp clock reaches a specified time.