Thermographic measurement techniques15
3. Measure the radiation intensity (= apparent temperature) from the reflection source us-
ing the following settings:
Emissivity: 1.0
D
obj
: 0
You can measure the radiation intensity using one of the following two methods:
Figure 15.3 1 = Reflection source Figure 15.4 1 = Reflection source
You can not use a thermocouple to measure reflected apparent temperature, because a
thermocouple measures temperature, but apparent temperatrure is radiation intensity.
15.2.1.1.2 Method 2: Reflector method
Follow this procedure:
1. Crumble up a large piece of aluminum foil.
2. Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size.
3. Put the piece of cardboard in front of the object you want to measure. Make sure that
the side with aluminum foil points to the camera.
4. Set the emissivity to 1.0.
#T559918; r. AN/42281/42281; en-US
61

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the FLIR C3 and is the answer not in the manual?

FLIR C3 Specifications

General IconGeneral
Temperature Range-10°C to 150°C (14°F to 302°F)
Field of View41° x 31°
Thermal Sensitivity<0.10°C
Accuracy±2 °C or ±2% of reading
Resolution80 x 60 pixels
Image StorageInternal memory
ConnectivityWi-Fi
Display3-inch touchscreen
Image ModesThermal, MSX, Picture-in-Picture
Digital Camera640 x 480 pixels
File FormatJPEG
Battery Life4 hours
Dimensions125 x 80 x 24 mm (4.9 x 3.1 x 0.94 in)
Operating Temperature-10°C to +50°C