Loading...
Achieving a good image14
Reflected appa-
rent temperature
+20°C (+69°F)
Relative humidity 50%
Atmospheric
temperature
+20°C (+69°F)
14.9.4 Procedure
Follow this procedure to change the object parameters globally:
1. Turn the mode wheel to
or .
2. Push the
button to display a menu.
3. Move the joystick left/right to go to the Edit tab.
4. Move the joystick up/down to select Object parameters.
5. Push the joystick to display a dialog box.
6. Move the joystick up/down to select the parameter you want to change, then push the
joystick.
7. Move the joystick up/down to change the value, then push the joystick.
8. Push the
button to confirm and leave the setup mode.
Note
Of the seven parameters above, emissivity and reflected apparent temperature are the
two most important to set correctly in the camera.
To change object parameters locally, first select a measurement tool in the toolbox, then
select Use local parameters. Change the local parameters by selecting Edit local pa-
rameters, then edit them in the same way as for global object parameters.
14.9.5 Related topics
For in-depth information about parameters, and how to correctly set emissivity and re-
flected apparent temperature, see 32 Thermographic measurement techniques.