EasyManua.ls Logo

FLIR Lepton 1.6 - Page 24

FLIR Lepton 1.6
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FLIR LEPTON® Engineering Datasheet
The information contained herein does not contain technology as defined by the EAR, 15 CFR 772, is publicly available,
and therefore, not subject to EAR. NSR (6/14/2018).
Information on this page is subject to change without notice.
Lepton Engineering Datasheet, Document Number: 500-0659-00-09 Rev: 203
24
Table 2 - Telemetry Data Content and Encoding
Telemetry
Row
Word
start
Word
End
Number
of 16-bit
Words
Name
Notes
A
0
0
1
Telemetry
Revision
Format = major (byte 1), minor rev (byte
0).
A
1
2
2
Time Counter
32-bit counter in units of msec elapsed
since
boot-up
A
3
4
2
Status Bits
See Table 3
A
5
12
8
Module serial #
A
13
16
4
Software revision
A
17
19
3
=
3
3
Reserved
A
20
21
2
Frame Counter
32-bit counter of output frames
A
22
22
1
Frame Mean
A
23
23
1
FPA Temp
In counts (prior to conversion to Kelvin)
A
24
24
1
FPA Temp
In Kelvin x 100
A
25
25
1
Housing Temp
In counts (prior to conversion to Kelvin)
Lepton 2.5, 3.5
A
26
26
1
Housing Temp
In Kelvin x 100
Lepton 2.5, 3.5
A
27
28
2
Reserved
A
29
29
1
FPA Temp at last
FFC
Updated every FFC. Units are Kelvin x100

Table of Contents