EasyManuals Logo

fluidra 66178 User Manual

fluidra 66178
372 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #19 background imageLoading...
Page #19 background image
0000137242 Rev. 7.1 17
3.3.2.2 BY REFERENCE MODE
Connect the ORP probe to the instrument as shown in the electrical connections section.
Select the ORP probe in the Calibration menu.
Select the second option (By Ref Calibration)
This type of calibration can be performed WITHOUT removing the probe from the probe holder, simply reading the
ORP value in order to make the appropriate correction on the read measure. If a standard calibration will be
performed, the value set in this By Ref calibration, will be canceled!
3.3.3 CL (CHLORINE) PROBE CALIBRATION
Connect the probe to the instrument as indicated in the electrical connections.
Select the CL probe from the Calibration menu.
2B1 period one
CL CAL. Type: MAN
0.50 ppm
CL CAL. Type: MAN
1.20 ppm
CL CAL. Type: MAN
1.20 ppm
Wait 10”
Use a reference instrument to read the
chlorine value.
Adjust the value shown on the display to
match the value read by the reference
instrument. Press Enter to confirm.
Wait 10 seconds for the calibration to
complete.
Once the operation has concluded, a
message will appear indicating that the
calibration has been carried out successfully.
2
-----------------
Calibration
----------------------
2B1 One Point
2B2 Two Points
2
----------------
Calibration
---------------------
+475mV
2
----------------
Calibration
---------------------
+500mV
Wait
On ByRef option:
Is displayed the reading of the ORP value,
without calibration and flashing
The value can be modified
Set the real ORP value
Confirm with Enter
Upon confirmation, the ORP value is
displayed, stops flashing and the underlying
tag “Wait” starts flashing.
After a few seconds the system
automatically returns to the previous menu
(selection of the calibration type)

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the fluidra 66178 and is the answer not in the manual?

fluidra 66178 Specifications

General IconGeneral
Brandfluidra
Model66178
CategoryController
LanguageEnglish