Table 6. Command Summary (cont.)
Keyword Description
:MUNits
PERCent|DB50|DB75|DB600|DB
Sets the Units for MATH calculations. Default +1.999999999e
PERCent = Units of (mx - c) in percent of z
dB50 =Relative to 1 mW in 50 Ω
dB75 =Relative to 1 mW in 75 Ω
dB600 =Relative to 1 mW in 600 Ω
dB = Relative to 1
:MUNits? Returns the MUNit parameter
:MUNits:STATe <boolean> Turns MUNits ON or OFF. Default OFF.
:MUNits:STATe? Returns the MUNits setting
:STATe <Boolean> Turn Math ON or OFF. Default OFF.
:STATe? Returns the Math state
:RESet Reset Math factors to defaults
:ZFACtor <nrf>|LREading Sets the z variable to a value or the last reading. Default 1, min -
9.9E+37, Max 9.9E+37.
:ZFACtor? Returns the z variable value
:ZFACtor:STATe <boolean> Turns the - z operation ON or OFF. Default OFF.
:ZFACtor:STATe? Returns the state of z operation (I = On, 0 = Off)
CALCulate:SSTatistics
:AVERage?
:COUNt? Returns the count of elements in memory/record in the statistics
calculation
:MAXimum? Returns the maximum value
:MINimum? Returns the minimum value
SDEViation [PPM] Sets the Standard Deviation value (no parameter) or Standard
Deviation in ppm
:SDEViation? Returns the standard deviation
:SPAN? Retuns the span of the readings
:STATe <boolean> Turns statistics state ON or OFF. Default OFF.
:STATe? Returns the statistics state