Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
21
Table 6. Command Summary (cont.)
Keyword Description
CALibration
CALibration
:COPY:BASeline Copy the certified stores over the Baseline stores
:SECure? Returns calibration adjustment mode "FACTory", "USER" or
"NONe"
:SECure:EXIT Exit from calibration adjustment mode
:SECure:PASSword <string> Enables calibration adjustment mode
:STORe:DATe? <CERTified|BASeline> Returns the date of the last calibration adjustment of the store
<year>,<month>,<day>,<Hours>,<Minutes>,<seconds>
:STORe:USE <CERTified|BASeline> Sets the active stores. Default CERTified.
:STORe:USE? Returns the active stores "CERTified" or "BASeline"
:TARget:DESCription? Returns the target name as a quoted string containing target
point description
:TARGet:INPut? Returns the target input field description as a quoted string
:TARGet:STEP <nrf> Sets the calibration adjust step
:TARGet:VALue <nrf> Sets the target value
:TARGet:VALue[?] Returns the target value
:TRIGger? Initiates the adjustment, returns 0 = success, 1 = failure
DISPlay Subsystem
DISPlay[:STATe]<boolean> Turn the display ON or OFF. Default ON.
DISPlay
:BRIGhtness <nrf>
Set the display brightness. Default 50 %
:BRIGhtness? :BRIGhtness? Retrieve Brightness setting
:DTIMe<nrf> Set the display backlight timer in seconds. Default 1800
seconds (30 minutes)
:DTIMe? Returns the time setting
DISPlay[:STATe]? Returns the state of the display