8588A/8558A
Remote Programmer's Manual
32
Table 6. Command Summary (cont.)
Keyword Description
:RANGe:AUTO[:STATe] <boolean> Sets Auto ranging ON or OFF
:RANGe:AUTO[:STATe]? Returns the Auto range setting
:RANGe <range>|MIN|MAX|DEF
Sets the expected maximum input resistance, or if selected, to
the MIN, MAX, or Default setting
:RANGe? [MIN|MAX|DEF]
Returns the range setting or if selected the MIN, MAX, or
Default setting
:RESolution <resolution>|MIN|MAX|DEF
Set maximum expected value or MIN, MAX, or Default
resolution
:RESolution? [MIN|MAX|DEF]
Returns the selected resolution, or if specified, the MIN, MAX, or
Default resolution
:FILTer[:STATe] <boolean> Turns the input filter ON or OFF
:FILTer[:STATe]? Returns the filter setting, 0 = OFF, 1 = ON
[:SENSe]:TEMPerature:RTD|TRTD|FRTD
(RTD= 2-wire TRTD = three wire, FRTD = 4-wire Resistance temperature device)
:APERture <seconds>|MIN|MAX|DEF
Sets the ADC aperture value in seconds or to the MIN, MAX,
or Default setting
:APERture? [MIN|MAX|DEF]
Returns the aperture setting or the MIN, MAX, or Default
settings in seconds
:APERture:MODE AUTO|FAST|MANual Sets the aperture mode
:APERture:MODE? Returns the aperture mode
:NPLC <plcs>|MIN|MAX|DEF
Sets the ADC aperture in number of power line cycles or to
the MIN, MAX, or Default plc setting
:NPLC? [MIN|MAX|DEF]
Returns the ADC aperture in number of power line cycles or
to the MIN, MAX, or Default plc setting
:RESistance:RO 25|100 Set the reference resistance value
:RESistance:RO? Return the reference resistance value
:RESolution <resolution>|MIN|MAX|DEF
Set maximum expected value or min, max or default
resolution. MAX or MIN sets resolution to maximum/minimum
appropriate to the NPLC/Aperture setting.
:RESolution? [MIN|MAX|DEF]
Returns the selected resolution, or if specified, the MIN,
MAX, or Default resolution
:SECondary[:STATe] <boolean> Turns the secondary reading (voltage) ON or OFF
:SECondary[:STATe]? Returns the state of the secondary reading (1 = ON, 0 = OFF
:UNITs C|CEL|F|FAR|K Set units to Celsius (C, CE), Fahrenheit (F, FAR) or Kelvin
:UNITs?