8588A/8558A
Remote Programmer's Manual
42
Table 7. SENSe Subsystem Power On and Reset Default Settings
DCV
DCV is the default function
Range: 1000 (Min 0.1, Max 1000)
Resolution: 0.0001 (7 digits) (min 4 digits, max 8
digits, actual value is range dependent)
Aperture: AUTO (default 1 second, Min 0, Max 10
seconds), NPLC Min = 0 Max = 600
Input impedance: AUTO
DCI
Range 30
Resolution 0.000001 (7 digits) (min 4 digits, max 7
digits, actual value is range dependent)
Aperture: AUTO (default 1 second, Min 0, Max 10
seconds), NPLC Min = 0 Max = 600
ACV
Range: 1000
Resolution: 0.0001 (7 digits) (min 4 digits, max 7
digits, actual value is range dependent)
RMS Filter: 40 Hz
Signal path coupling/impedance: AC, 1MΩ
Frequency path coupling: AC (Disabled)
Frequency path bandwidth limit: OFF
Secondary reading: Frequency
Counter Gate: Auto
Bandwidth Limit: Wideband
Peak to Peak Method: Measured
ACI
Range: 30 A
Resolution:0.00001 (6 Digits) (min 4 digits, max 7
digits, actual value is range dependent)
RMS Filter: 40 Hz
Signal Path Coupling: AC
Secondary reading: Frequency
Frequency path coupling: AC (Disabled)
Frequency path bandwidth limit: OFF
Counter Gate Time: Auto
Peak to Peak Method: Measured (Disabled)