Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
43
Table 8. SENSe Subsystem Power On and Reset Default Settings (cont.)
Resistance 2W Normal
Range: 10000 (10 kΩ)
Resolution: 0.01 (7 digits) (min 4 digits, max 8 digits,
actual value is range dependent)
LOI: OFF
Aperture: AUTO (default 1 second, Min 100 µs, Max
10 seconds), NPLC Min = 0 Max = 600
Filter: OFF
Resistance 4W Normal
Range: 10000 (10 kΩ)
Resolution: 0.01 (7 digits) (min 4 digits, max 8 digits,
actual value is range dependent)
LOI: OFF
Aperture: AUTO (default 1 second, Min 100 µs, Max
10 seconds), NPLC Min = 0 Max = 600
Filter: OFF
Resistance 4W Tru
Range: 10000 (10 kΩ)
Resolution: 0.01 (7 digits) (min 4 digits, max 8 digits,
actual value is range dependent)
LoI: OFF
Aperture: AUTO (default 1 second, Min 100 µs, Max
10 seconds), NPLC Min = 0 Max = 600
Filter: OFF
Resistance 2W HV
Range: 10000000 (10 MΩ)
Resolution: 1.0 (7 digits) (min 4 digits, max 8 digits,
actual value is range dependent)
Aperture: AUTO (default 1 second, Min 100 µs, Max
10 seconds), NPLC Min = 0 Max = 600
Filter: OFF
Resistance 4W HV
Range: 10000000 (10 MΩ)
Resolution: 1.0 (7 digits)
Aperture: AUTO (default 1 second, Min 100 µs, Max
10 seconds), NPLC Min = 0 Max = 600
Filter: OFF
Capacitance
Range: 1 mF
Resolution: 5 digits (min 4 digits, max 5 digits, actual
value is range dependent)
LoI OFF