Reference Multimeter and 8 ½ Digit Multimeter
SCPI Commands Reference
45
Table 8. SENSe Subsystem Power On and Reset Default Settings (cont.)
Thermocouple
Type: K
Second Reading: ON
Resolution: Default = 0.1 (5 Digits) Max = 0.01
(6 Digits)
Units: Celsius
Aperture: Auto, (default 100 ms, Min 0, Max 10
seconds)
Digitize
V or I: Volts
Range: 1000
Coupling/Zin: DC, Auto
Aperture: default 0, Min 0, Max 0.003 seconds
Low Pass Filter: 3MHz
RF Power
Reading: Absolute
Average: Auto
Units: dBm
Reference Frequency: 0.0 Hz
Reference Level: -99.0 dBm