106/107
Users Manual
30
Function Range Resolution
Accuracy
106 107
Frequency
[2]
Hz
(10 Hz to 100 kHz)
50.00 Hz
500.0 Hz
5.000 kHz
50.00 kHz
100.0 kHz
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
0.1 kHz
NA 0.1 % + 3
Duty Cycle
[2]
1 % to 99 % 0.1 % NA 1 % typical
[3]
[1] Specifications do not include errors due to test lead capacitance and capacitance floor (may be up to 1.5 nF in the
50 nF range).
[2] All AC, Hz, and duty cycle readings are specified from 1 % to 100 % of range. Inputs below 1 % of range are not
specified.
[3] Typical means when the frequency is at 50 Hz or 60 Hz and the duty cycle is between 10 % and 90 %.