True-rms Multimeters 
 Power-Up Options  
 9 
When you turn the Meter on, it defaults to Autorange and 
Auto is displayed. 
1. To enter the Manual Range mode, press  
q. 
Manual is displayed. 
2. In the Manual Range mode, press 
qto increment 
the range. After the highest range, the Meter wraps to 
the lowest range. 
Note 
You cannot manually change the range in the 
MIN MAX AVG or Display HOLD modes.  
If you press qwhile in MIN MAX AVG or
 
Display Hold
, the Meter beeps twice, indicating 
an invalid operation and the range does not 
change. 
3. To exit Manual Range, press 
q for at least 1 second 
or turn the rotary switch. The Meter returns to 
Autorange and Auto is displayed. 
Power-Up Options 
To select a Power-Up Option, hold down the button 
indicated in the following table while turning the Meter on. 
Power-Up Options are canceled when you turn the Meter 
off and when sleep mode is activated. 
 
Button  Power-Up Options  
f 
Turns on all display segments.  
p 
Disables beeper. bEEP is displayed when 
enabled. 
q 
Enables low impedance capacitance 
measurements. LCAP is displayed when 
enabled. See page 14. 
g 
Disables automatic power-down ("Sleep 
mode"). PoFF is displayed when enabled. 
Q 
Disables auto backlight off. LoFF is 
displayed when enabled. 
Making Basic Measurements 
The figures on the following pages show how to make 
basic measurements.  
When connecting the test leads to the circuit or device, 
connect the common (COM) test lead before connecting 
the live lead; when removing the test leads, remove the live 
lead before removing the common test lead. 