7 Digital Communications Interface
Remote Commands
Command Description
Measurement Commands
T read measurement (includes label, unit, and time)
F[ETCH?] read measurement value (SCPI compatible)
SA[=[[[hh:]mm:]ss] read [or set] serial sample period
U=C|F|K|O select units
ST[=ON/OF] read [or set] the time stamp
CL[=hh:mm:ss] read [or set] the system clock
Probe Characterization Commands
PR[=90|68|R|S] read [or select] the characterization type
R0[=<value>] read [or set] R0 or R(0.01°C) depending on the selected characterization
AL[=<value>]
read [or set]
α
for the Callendar-Van Dusen or IPTS-68 characterization
DE[=<value>]
read [or set]
δ
for the Callendar-Van Dusen or IPTS-68 characterization
BE[=<value>]
red [or set]
β
for the Callendar-Van Dusen characterization
A4[=<value>] read [or set]
a
4
for the ITS-90 or IPTS-68 characterization
B4[=<value>] read [or set]
b
4
for the ITS-90 characterization
C4[=<value>] read [or set]
c
4
for the IPTS-68 characterization
A
n
[=<value>] read [or set]
a
6
,a
7
,a
8
,a
9
,a
10
,or
a
11
for the ITS-90 characterization
B
n
[=<value>] read [or set]
b
6
,b
7
,b
8
,
or
b
9
for the ITS-90 characterization
C
n
[=<value>] read [or set]
c
6
or
c
7
for the ITS-90 characterization
D6[=<value>] read [or set]
d
6
for the ITS-90 characterization
SC[=68/90] read [or set] IPTS-68 scale conversion
CO=<value> test resistance to temperature conversion
Sample Parameter Commands
FI[=<value>] read [or set] filter time constant
CU[=<value>] read [or set] probe current
PS[=<value.] read [or set] the power saver period
Communication parameter commands
DU[=F/H] read [or set] serial sample duplex mode
LF[=ON/OF] read [or set] serial linefeed
Table 6 Command List