The following tables can be used to determine the maximum or minimum display values considering maximum instrument
operating error per EN61557-1, 5.2.4.
Insulation Resistance Maximum and Minimum Display Values
50 V 100 V 250 V 500 V 1000 V
Limit
Value
Minimum
Display
Value
Limit
Value
Minimum
Display
Value
Limit
Value
Minimum
Display
Value
Limit
Value
Minimum
Display
Value
Limit
Value
Minimum
Display
Value
0.05 0.07 0.05 0.07 0.05 0.07 0.05 0.07
0.06 0.08 0.06 0.08 0.06 0.08 0.06 0.08
0.07 0.09 0.07 0.09 0.07 0.09 0.07 0.09
0.08 0.10 0.08 0.10 0.08 0.10 0.08 0.10
0.09 0.12 0.09 0.12 0.09 0.12 0.09 0.12
0.1 0.13 0.1 0.13 0.1 0.13 0.1 0.13 0.1 0.1
0.2 0.26 0.2 0.26 0.2 0.26 0.2 0.26 0.2 0.3
0.3 0.39 0.3 0.39 0.3 0.39 0.3 0.39 0.3 0.4
0.4 0.52 0.4 0.52 0.4 0.52 0.4 0.52 0.4 0.5
0.5 0.65 0.5 0.65 0.5 0.65 0.5 0.65 0.5 0.7