1587 FC/1587/1577 
Users Manual 
24 
Insulation 
Insulation tests should only be performed on dead circuits. 
Check the fuse before testing. See Fuse Test later in this 
manual. To measure insulation resistance, set up the Meter 
as shown in Figure 9 and follow the steps below: 
1.  Insert test probes in the 
p and q input terminals. 
2.  Turn the knob to 
INSULATION position. A battery load 
check is initiated when the switch is moved to this 
position. If the battery fails the test B and bat appear 
in the lower display. Insulation tests cannot be 
performed until the batteries are replaced. 
3. Press r to select the voltage. 
4.  Connect the probes to the circuit to be measured. The 
Meter automatically detects if the circuit is energized. 
•  The primary display shows - - - - until you press 
s and a valid insulation resistance reading is 
obtained. 
•  The high voltage symbol (Z) along with a primary 
display of >30 V warns if voltage more than 30 V 
ac or dc is present. In this condition, the test is 
inhibited. Disconnect the Meter and remove power 
before proceeding. 
5.  Push and hold s to start the test. The secondary 
display shows the test voltage applied to the circuit 
under test. The high voltage symbol (Z) along with a 
primary display showing the resistance in MΩ or GΩ 
appears. The T icon appears on the lower portion of 
the display until s is released.  
When resistance is higher than the maximum display 
range, the Meter displays the Q symbol and the 
maximum resistance for the range. 
6.  Keep the probes on the test points and release the 
s button. The circuit under test then discharges 
through the Meter. The resistance reading remains on 
the primary display until a new test is started, a 
different function/range is selected, or >30 V is 
detected.