Energy Logger
Electrical Specifications
69
Influence of conductor position
in jaw opening
.........................................................±0.5 % of reading (@ 50/60 Hz)
Bandwidth ...............................................................10 Hz to 2.5 kHz
Working Voltage......................................................600 V CAT III, 300 V CAT IV
[1] Reference Condition:
Environmental: 23 °C ±5 °C, no external electrical/magnetic field, RH 65 %
Primary conductor in center position
Size (H x W x L) ......................................................110 mm x 50 mm x 26 mm
(4.33 in x 1.97 in x 1.02 in)
Maximum conductor size ........................................15 mm (0.59 in)
Output cable length.................................................2 m (6.6 ft)
Weight.....................................................................190 g (6.70 oz)
Material ...................................................................Case ABS and PC
Output cable: TPR/PVC
Temperature operating............................................-10 °C to +55 °C
(-14 °F to 131 °F)
Temperature, non-operating ...................................-20 °C to +70 °C
(-4 °F to 158 °F)
Relative Humidity, operating ...................................15 % to 85 % non-condensing
Max Operating Altitude ...........................................2000 m (6500 ft)
up to 4000 m (13 000 ft) derate to 600 V CAT II/300 V CAT IV
Max Storage Altitude ..............................................12 km (40 000 ft)
Warranty..................................................................1 year