5. Insert the sensor cable through the strain relief .
The cable diameter must be in the range of 3 mm to 6 mm to
ensure proper strain relief.
6. Tighten the strain relief.
7. To connect the wires to the terminals E, press the opening
lever.
8. Correctly orient the lid and close the housing.
9. Tighten the captive screws .
Note
Pay attention to the correct orientation of the lid.
10. Connect the 4-pin connector (not shown) to the Logger.
11. Configure the AUX channel on the Logger to show the
correct readings and unit of measurement for the attached
sensor. For more information, see the documentation for
the Logger and the Sensor.
Voltage Divider Input
1 V DC to 1000 V DC Configuration
To connect the Adapter:
1. Use correctly rated 4 mm test leads on the safety socket
inputs.
Note
Do not connect a signal at the direct input and voltage
divider input at the same time. The measured value will
be incorrect.
2. Connect the 4-pin connector to the Logger.
3. Configure the AUX channel on the Logger to show the
correct readings, select an available custom configuration
(Custom 1 to Custom 5), and use these settings:
Sensor: 0-10 V
Unit: V
Minimum: 0 V
Maximum: 1000 V
Shop for Power Metering products online at:
1.888.610.7664
www.PowerMeterStore. com

Other manuals for Fluke 1736/FPC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1736/FPC and is the answer not in the manual?

Fluke 1736/FPC Specifications

General IconGeneral
BrandFluke
Model1736/FPC
CategoryData Loggers
LanguageEnglish

Related product manuals