175, 177, & 179
Calibration Information
10
Performance Tests
WWarning
To avoid electric shock, do not perform the performance test procedures unless
the Meter is fully assembled.
The following performance tests verify the complete operation of the Meter and check the accuracy of
each meter function against the Meter’s specifications. If the Meter fails any part of the test, calibration
adjustment and/or repair is indicated.
In the performance tests, the Meter is referred to as the unit under test (UUT).
Required Equipment
A Fluke 5500A Multi-Product Calibrator (or equivalent) is required for the performance test procedures
in this document.
If an equivalent calibrator is used, it must meet the accuracy specifications shown in Table 1.
Table 1. Calibrator Specifications
Recommended Equipment Measurement Function Accuracy
5500A Multi-Product Calibrator
(or equivalent)
DC Volts 0 to 1000 V
± 0.075 %
DC Current 0 to 10 A
± 0.425 %
AC Volts 0 to 1000 V
± 0.375 % @ 45 Hz to 1 kHz
AC Current 0 to 10 A
± 0.55 %
Resistance 0 to 4 M
± 0.125 %
4 to 40 M
± 0.5 %
Capacitance 500 pF to 5 µF
± 0.325 %
Temperature .25%
Frequency 0.5 Hz to 100 kHz
± 0.005 %
Fluke 80 AK
Thermocouple Adapter Accessory
K-type -
K-type Thermocouple K-type, mini-plug on both ends -

Other manuals for Fluke 179

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 179 and is the answer not in the manual?

Fluke 179 Specifications

General IconGeneral
Frequency range0 - 100 kHz
DC current range0.01 - 10 A
DC voltage range0.1 - 1000 V
Resistance range0.1 - 50000000 Ω
Capacitance range1 - 10000 nF
Resistance resolution0.1 Ω
Resolution (frequency)0.01 Hz
Resolution (DC current)0.01 µA
Resolution (DC voltage)0.1 mV
Resolution (temperature)0.1 °C
Basic accuracy (frequency)±(0.1% + 1)
Basic accuracy (DC voltage)±(0.09% + 2)
Basic accuracy (resistance)±(0.9% + 1)
Basic accuracy (capacitance)±(1.2% + 2)
Temperature measurement range-40 - 400 °C
Storage temperature (T-T)-30 - 60 °C
Operating temperature (T-T)-10 - 50 °C
Operating relative humidity (H-H)0 - 90 %
ManualYes
Weight and Dimensions IconWeight and Dimensions
Depth90 mm
Width43 mm
Height185 mm
Weight420 g

Related product manuals