Table 2. Performance Tests 175 (continued)
Meter Response
Test
(Switch
Position)
5520 Output Lower Limit Upper Limit
3 V 2.993 V DC 3.007 V DC
300 V 299.3 V DC 300.7 V DC
1000 V 996 V DC 1004 V DC
L
DC Volts
1000 V 1004 V DC 996 V DC
1 V 45 Hz 44.94 Hz 45.06 Hz
1 V 2 kHz 1.997 kHz 2.003 kHz
1.5 V 20 kHz 19.97 kHz 20.03 kHz
* Hz
Frequency
5 V 50 kHz 49.94 kHz 50.06 kHz
0.03 V 0 Hz 29.8 mV DC 30.2 mV DC
-0.3 V 300.7 mV DC 299.3 mV DC
0.6 V 598.9 mV DC 601.1 mV DC
1.5 V 20 kHz 19.97 kHz 20.03 kHz
E
Millivolts DC
5 V 50 kHz 49.94 kHz 50.06 kHz
25 Beeper OnR
Continuity
250 Beeper Off
19 18.6 19.4 e
Ohms
2
19000000 18.68 M 19.32 M
G
Diode
1
Apply 2.0 V 1.978 V DC 2.022 V DC
E
Capacitance
1
Apply 0.90 uF 887 nF 913 nF
4.0 A 45 Hz 3.937 A AC 4.063 A AC?
Amps AC
9.0 A 1 kHz 8.83 A AC 9.17 A AC
0.003 A 45 Hz 2.92 mA AC 3.08 mA AC
0.05 A 1 kHz 49.22 mA AC 50.78 mA AC
F
AC Milliamps
0.4 A 1 kHz 393.7 mA AC 406.3 mA AC
0.003 A, 0 Hz 2.94 mA DC 3.06 mA DC
0.05 A 49.47 mA DC 50.53 mA DC
mI
DC Milliamp
1
0.4 A 404.3 mA DC 395.7 mA DC
4.0 A 3.957 A DC 4.043 A DCA
DC Amps
1
9.0 A 9.12 A DC 8.88 A DC
1. Press the YELLOW button to access this function.
2. Does not include test lead resistance.

Other manuals for Fluke 179

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 179 and is the answer not in the manual?

Fluke 179 Specifications

General IconGeneral
Frequency range0 - 100 kHz
DC current range0.01 - 10 A
DC voltage range0.1 - 1000 V
Resistance range0.1 - 50000000 Ω
Capacitance range1 - 10000 nF
Resistance resolution0.1 Ω
Resolution (frequency)0.01 Hz
Resolution (DC current)0.01 µA
Resolution (DC voltage)0.1 mV
Resolution (temperature)0.1 °C
Basic accuracy (frequency)±(0.1% + 1)
Basic accuracy (DC voltage)±(0.09% + 2)
Basic accuracy (resistance)±(0.9% + 1)
Basic accuracy (capacitance)±(1.2% + 2)
Temperature measurement range-40 - 400 °C
Storage temperature (T-T)-30 - 60 °C
Operating temperature (T-T)-10 - 50 °C
Operating relative humidity (H-H)0 - 90 %
ManualYes
Weight and Dimensions IconWeight and Dimensions
Depth90 mm
Width43 mm
Height185 mm
Weight420 g