175, 177, & 179
Calibration Information
16
Table 5. Calibration Steps
Function
(Switch Position) Calibration Step Input Value
C-01 6.000 V, 900.0 Hz
C-02
600.0 mV, 100 Hz
C
(AC Volts)
C-03 6.000 V, 100 Hz
C-04 6.000 V
C-05 60.00 V
C-06 600.0 V
L
(DC Volts)
C-07 1000 V
C-08
600.0 mV
T E
DC Millivolts
(temperature - 179 only)*
C-09 0.0 °C
e
(Ohms)
C-09 (175/177)
C-10 (179)
6.000 k
R
(Continuity)
C-10 (175/177)
C-11 (179)
600
mA
(Milliamps)
C-11 (175/177)
C-12 (179)
600.0 mA dc
A
(Amps)
C-12 (175/177)
C-13 (179)
6.000 A dc
* To ensure accuracy, the Meter and thermocouple adapter must be at the same temperature. Allow 20 minutes before performing the
temperature procedures
.