2620A, 2625A
Users Manual
4-42
Table 4-8. Command and Query Reference (cont)
RWLS Remote with Lockout (RS-232 only)
Enter the IEEE-488.1 remote with front panel lockout (RWLS) state. All
front panel buttons are disabled, and the REM annunciator is lit.
If this command is used with the IEEE-488 interface, an Execution Error
is generated.
SCALE_MB Set Mx+B Scaling Values
Set the M and B scaling values for the indicated channel, and display the
results of the Mx+B calculation in the indicated display range. Successful
execution of this command clears all
values in the Review array (all channels.)
SCALE_MB <channel>,<M_value>,<B_value>,<disp_range>
<channel> = (0 .. 20)
<M_value> = signed numeric quantity
<B_value> = signed numeric quan<disp_range> = (16)
<disp_range> = (1 .. 16)
CODE DISPLAY MAX OFFSET CODE DISPLAY MAX
OFFSET
VALUE VALUE
1 0.0000 m 0.01 9 0.0000 k 1.0E4
2 00.000 m 0.1 10 00.000 k 1.0E5
3 000.00 m 1.0 11 000.00 k 1.0E6
4 0000.0 m 1.0 12 0000.0 k 1.0E6
5 0.0000 x1 10.0 13 0.0000 M 1.0E7
6 00.000 x1 100.0 14 00.000 M 1.0E8
7 000.00 x1 1000.0 15 000.00 M 1.0E9
8 0000.0 x1 1000.0 16 0000.0 M 1.0E10
When M=1 and B=0, Mx+B scaling is effectively nonexistent. The values
for M and B must be in the span +/-9999.9 Mega (E+6).
An Execution Error is generated for any of the following:
Invalid entries for channel number, M or B values, or display range code
are used.
The range required by the B value is larger than allowed by the display
range code.
The channel is defined as OFF.
Measurements are active.
Mx+B scaling values for a channel are automatically reset to 1 (M) and 0
(B) when the function for that channel is changed.