EasyManua.ls Logo

Fluke 2620A - Thermocouple Temperature

Fluke 2620A
197 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2620A, 2625A
Users Manual
3-8
Table 3-5. Thermocouple Temperature
Channel Function
Range
(Note)
PRESS
THESE
BUTTONS:
G
D
F
G
D
E
G
D
E
TO
SELECT
FROM
THESE
CHOICES:
1
.
.
20
OFF
V DC
V AC
Hz
°C or °F
J
K
E
T
N
R
S
b
C
Pt
(Completes
Selection and
returns to
Inactive Mode)
Note: The nine thermocouple choices and related temperature measurement ranges are:
"J"
"K"
"E"
"T"
"N"
"R"
"S"
"b"
"C"
Type J (-210 to 760
°
C)
Type K (-270 to 1372
°
C)
Type E (-270 to 1000
°
C)
Type T (-270 to 400
°
C)
Type N (-270 to 1300
°
C)
Type R (0 to 1767
°
C)
Type S (0 to 1767
°
C)
Type B (0 to 1820
°
C)
Type C (0 to 2316
°
C)*
(Tungsten - 5% Rhenium vs. Tungsten - 26% Rhenium)
"Pt" selects RTD temperature measurement (DIN/IEC 751). See Table 3-6.
* Hoskins Engineering Co.