EasyManua.ls Logo

Fluke 2620A - Totalizing and Digital Input Specifications

Fluke 2620A
197 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2620A/2625A
Users Manual
A-12
Maximum Autoranging Time
(Seconds per Channel)
Function Range Change Slow Fast
VDC 300 mV ----------------------------> 150 V 0.25 0.19
150 V 300 mV ----------------------------> 0.26 0.19
VAC 300 mV ----------------------------> 150V 4.50 4.12
150 V 300 mV ----------------------------> 1.38 1.08
Ohms 300 ----------------------------> 10.0 M 1.30 0.59
10.0 M 300 ----------------------------> 1.81 0.75
Totalizing Input
Input Voltage
30V maximum
4V minimum
2V peak minimum signal
Isolation
None
dc-coupled
Threshold
1.4V
Hysteresis
500 mV
Input Debouncing
None or 1.66 ms
Rate
0 to 5 kHz with debouncing off
Maximum Count
65,535
Digital Inputs
Input Voltage
30V maximum
-4V minimum

Table of Contents

Related product manuals