EasyManua.ls Logo

Fluke 2620A - Sample Program

Fluke 2620A
197 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2620A, 2625A
Users Manual
4-10
10 ’ EXAMPLE.BAS Hydra program to scan VDC, VAC, OHMS, FREQ, or TEMP
20 ’ - initialize RS232 communication and set up Hydra
30 ’ - display and record measurement data in "TESTDATA.PRN"
40 ’ Hydra must be set up for RS232, 9600 baud, no parity (from front panel)
50 KEY OFF
60 ’ Open communications port 9600 baud, no parity, 8 bit data,
70 ’ ignor Clear to Send, Data Set Ready, Carrier Detect
80 OPEN "COM1:9600,N,8,,cs,ds,cd" FOR RANDOM AS #1
90 IF ERRORCODE <> 0 THEN PRINT "ERROR - Could not open com1:" : END
100 ’
110 OPEN "testdata.prn" FOR OUTPUT AS #2 ’Open data file
120 IF ERRORCODE <> 0 THEN PRINT "ERROR - Could not open data file" : END
130 ’
140 PRINT #1, "ECHO 0" ’Turn off command echo
150 ’
160 NUMBEROFCHANS = 0
170 WHILE (NUMBEROFCHANS < 1) OR (NUMBEROFCHANS > 20)
180 INPUT "Enter number of channels (1-20): ", NUMBEROFCHANS
190 WEND
200 PRINT "(Wait...)"
210 FOR I = (NUMBEROFCHANS + 1) TO 20 ’ Turn unused channels off
220 PRINT #1, "FUNC" + STR$(I) + ",OFF"
230 GOSUB 800
240 NEXT I
250 ’
260 CLS
270 LOCATE 2,25 : PRINT "Sample Program for Hydra"
275 PRINT #1, "*IDN?" : GOSUB 800 : LINE INPUT #1, RESULT$
276 LOCATE 3,20 : PRINT RESULT$;
280 ’
290 WHILE (1)
300 ’ Print banner line at bottom of screen
310 LOCATE 25,1
320 PRINT "1 = VDC 2 = VAC 3 = OHMS 4 = FREQ 5 = TEMP 6 = QUIT";
330 ’
340 FUNC$ = "0"
350 WHILE (FUNC$ < "1") OR (FUNC$ > "6")
360 LOCATE 23,1 : INPUT " Selection: ", FUNC$
370 WEND
380 ’ Exit and clean up if "Quit"
390 If FUNC$ = "6" THEN CLOSE 1, 2 : CLS : KEY ON : END
400 ’
410 ’ Set up later part of command string to Hydra (function and range)
420 IF FUNC$ = "1" THEN CMD$ = "VDC, 1"
430 IF FUNC$ = "2" THEN CMD$ = "VAC, 1"
440 IF FUNC$ = "3" THEN CMD$ = "OHMS, 1, 2"
450 IF FUNC$ = "4" THEN CMD$ = "FREQ, 1"
460 IF FUNC$ = "5" THEN CMD$ = "TEMP, K"
470 ’
480 ’ Set up Hydra
485 LOCATE 23,1 : PRINT "Programming... ";
490 FOR I = 1 TO NUMBEROFCHANS ’ Program channels
Figure 4-2. Sample Program