Fluke 2620A - RTD Input Specifications

Fluke 2620A
197 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendices
Specifications
A
A-5
Input Impedance
100 M minimum in parallel with 150 pF maximum
Common Mode and Normal Mode Rejection
See Specifications, DC Voltage Inputs
Cross-Talk Rejection
Refer to Appendix D
Open Thermocouple Detect
Small ac signal injection and detection scheme before each measurement detects greater
than 1 to 4 k as open. Performed on each channel unless defeated by computer
command.
RTD Inputs
Type
DIN/IEC 751, 100e Platinum (385)
IEC 751, Amendment 1, 100 e Platinum (IPTS-68)
RTD 4 Wire Accuracy* (±°C)
Temperature Resolution 18°C to 28°C 0°C to 60°C
(°C) Slow Fast 90 Day
Slow
1 Year
Slow
1 Year
Fast
1 Year
Slow
1 Year
Fast
-200.00 0.02 0.1 0.06 0.06 0.48 0.07 0.49
0.00 0.02 0.1 0.09 0.09 0.55 0.13 0.59
100.00 0.02 0.1 0.10 0.11 0.58 0.17 0.64
300.00 0.02 0.1 0.14 0.14 0.65 0.24 0.75
600.00 0.02 0.1 0.19 0.20 0.76 0.37 0.93
* Sensor inaccuracies are not included
IEC 751, Amendment 1, 100 e Platinum (ITS-90)
RTD 4 Wire Accuracy* (±°C)
Temperature Resolution 18°C to 28°C 0°C to 60°C
(°C) Slow Fast 90 Day
Slow
1 Year
Slow
1 Year
Fast
1 Year
Slow
1 Year
Fast
-200.00 0.02 0.1 0.06 0.06 0.48 0.07 0.49
0.00 0.02 0.1 0.09 0.09 0.55 0.13 0.59
100.00 0.02 0.1 0.13 0.13 0.60 0.19 0.67
300.00 0.02 0.1 0.17 0.18 0.69 0.28 0.79
600.00 0.02 0.1 0.30 0.31 0.87 0.48 1.04
* Sensor inaccuracies are not included