EasyManua.ls Logo

Fluke 2620A - Ohms Input Specifications

Fluke 2620A
197 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendices
Specifications
A
A-9
Ohms Inputs
Range Resolution
Typical Full
Maximum
Current
Through
Maximum Open
Slow Fast
Scale Voltage Unknown Circuit Voltage
300 10 m 0.1 0.22 V 1 mA 3.2V
3 k 0.1 1 0.25V 110 µA 1.5 V
30 k 1 10 0.29 V 13 µA 1.5 V
300 k 10 100 0.68 V 3.2 .µA 3.2 V
3 M 100 1 k 2.25 V 3.2 µA 3.2 V
10 M 1 k 10 k 2.72 V 3.2 µA 3.2 V
4-Wire Accuracy ±(%±)
Range 18°C to 28°C 0°C to 60°C
90 Days, Slow 1 Year, Slow 1 Year, Fast 1 Year, Slow 1 Year, Fast
300 0.013% + 20 m 0.014% + 20 m 0.014% + 0.2 0.031% + 20 m 0.031% + 0.2
3 k 0.015% + 0.2 0.016% + 0.2 0.016% + 2 0.039% + 0.2 0.039% + 2
30 k 0.013% + 2 0.014% + 2 0.014% + 20 0.039% + 2 0.039% + 20
300 k 0.020% + 20 0.021% + 20 0.021% + 200 0.050% + 20 0.050% + 200
3 M 0.059% + 200 0.063% + 200 0.063% + 2 k 0.231% + 200 0.231% + 2 k
10 M 0.168% + 2 k 0.169% + 2 k 0.709% + 20 k 0.573% + 2 k 0.923% + 20 k
2-Wire Accuracy
Add 4.0 to accuracy specifications for channels 1 to 20, and add 20 m for channel 0.
Lead wire resistances are not included.
Input Protection
300V dc or ac rms on all ranges
Cross-Talk Rejection
Refer to Appendix D.

Table of Contents

Related product manuals