LCD Annunciators 
0 
nS 
Conductance Range Annunciator 
(nS): 
Top range of the resistance function is the conductance range. Displays 
conductance in nS (nanosiemens). 1000/nS converts to Mfl 
. 
(Example: 
2 nS converts to 500BAO.) Use for measuring resistance above 
32AAn . 
Select n , open test leads, press RANGE button twice. (Refer to item 
7 
for manual range operation.) 
 
. 1. Analog Bar Graph Display: 
Analog representation of input. Composed of 31 segments which illuminate 
starting from the left as the input increases. (See display inside front 
cover.) A minus sign (INA
■
) is displayed for reverse-polarity inputs. 
Updated 25 times per second. 
C) I • I I I 
3 
30 
300 
Decimal Point/Range Indicator: 
Decimal point position and the digits (3, 30, 300) under the decimal 
point indicate the range in use. 
Manual Range Annunciator: 
Displayed in the Manual Range mode or if the selected function has only 
one range. Absence of the indicator implies autorange mode in use. The 
meter powers-up in autorange. In autorange, the meter automatically 
selects the measurement range. Refer to item 
7 
for operation. 
0 
n 
Low Battery Annunciator: 
At least 40 hours of battery life remain when first displayed. Battery 
voltage is tested each time the function switch is moved to a new 
position. 
Negative Polarity Annunciator: 
Automatically indicates negative inputs. 
o
t. 
Overload Indication: 
Indicates that the input is too large for the input circuitry. (The 
location of the decimal point depends on the measurement range.) 
® 
OF 
Overflow Indication: 
Indicates that the calculated difference in the Relative mode is too 
large to display 
(>3999 
counts) and that the input is not overloaded. 
q
)Beeper (not illustrated): 
The beeper emits beeps, clicks, or a continuous tone. It is used for 
audible indication in the diode test mode, when operating the push 
buttons, and when a new reading is displayed in the Touch-Hold mode.