27
Users Manual
6
J HOLD H
Touch Hold
Mode Pushbutton
Press momentarily to enter Touch Hold
mode. In Touch
Hold
, the meter captures a stable measurement and
holds it in the display. The operator can watch the probes
while taking measurements in difficult or hazardous
circuits, then look at the display when convenient. The
meter beeps and the display is automatically updated
each time a new, stable measurement is made. Press
momentarily to manually update reading. Press and hold
for 2 seconds to exit Touch Hold
mode. If HOLD is
depressed (>1 second) while the function switch is moved
from OFF to any ON position, the Touch Hold
mode will
only update to a new reading when the HOLD button is
pressed and automatic Touch Hold
updates are
defeated. This is useful when you want to take a reading
at a specific time and hold it.
K MIN
Minimum Annunciator
Indicates that the meter is in the MIN/MAX recording
mode, and the value displayed is the minimum digital
reading taken since reset or since entering MIN/MAX.
Refer to item 9 for operation.
L MAX
Maximum Annunciator
Indicates that the meter is in the MIN/MAX recording
mode, and the value displayed is the maximum digital
reading taken since reset or since entering MIN/MAX.
Refer to item 9 for operation.
M L
Relative Annunciator
Indicates that the meter is in the Relative mode and that
the value displayed is relative (the difference between the
present measurement and the previously stored reading).
Refer to item 8 for operation.
N H
Touch Hold
Mode Annunciator
Displayed when the Touch Hold
mode is in use. Refer to
item 10 for operation.
O Mke
Resistance Annunciators
The appropriate annunciator (e, k, or M) is displayed for
the resistance range in use.
P nS
Conductance Range Annunciator (nS)
Top range of the resistance function is the conductance
range. Displays conductance in nS (nanosiemens). 1000