279 FC
Calibration Manual
12
Table 3. Performance Tests
Test
(Control Knob Position)
Input Lower Limit Upper Limit Units
Volts AC
5 V, 45 Hz 4.947 5.053
V ac
5 V, 500 Hz 4.247 5.753
3 V, 45 Hz 2.967 3.033
50 V, 45 Hz 49.47 50.53
50 V, 500 Hz 42.47 57.53
30 V, 45 Hz 29.67 30.33
500 V, 45 Hz 494.7 505.3
500 V, 500 Hz 424.7 575.3
1000 V, 45 Hz 987 1013
Volts AC, Frequency
1 V, 900 Hz 899.0 901.0
Hz
5 V, 50 Hz 49.94 50.06
DC Volts
0.01 V 0.008 0.012
V dc
-5 V -5.006 -4.994
50 V 49.94 50.06
10 V 9.97 10.03
-500 V -500.6 -499.4
1000 V 997 1003
DC Millivolts
3 mV 2.8 3.2
mV dc500 mV 499.4 500.6
-500 mV -500.6 -499.4

AC Millivolts
30 mV, 60 Hz 29.4 30.6
mV ac500 mV, 60 Hz 494.7 505.3
500 mV, 500 Hz 494.7 505.3
Ohms
6 5.8 6.2
500 497.3 502.7
0.6 k 0.596 0.604
k
5 k 4.974 5.026
50 k 49.74 50.26
500 k 497.4 502.6
5 M 4.974 5.026
M
50 M 49.22 50.78
Capacitance
10 nF 8 12
nF
900 nF 887 913
9 µF8.879.13
µF90 µF 88.7 91.3
900 µF 887 913
Continuity
24 Beeper On
251 Beeper Off
Diode
1.95 V 1.928 1.972 V dc
Simulation
3.6 mV, 60 Hz 96.5 103.5
A ac36 mV, 60 Hz 965.0 1035
120 mV, 100 Hz 1940 2060
IR CAMERA
-5 °C-100
°C50 °C+45+55
180 °C +171 +189