True-rms Thermal Multimeter
Before Calibration Adjustment
13
Before Calibration Adjustment
Before you adjust the calibration, go to the Multimeter Calibration menu:
1. Push  to open the Setup Menu. 
2. Use  to highlight Calibration. 
3. Push   (Select) to open the submenu.
To enter the password:
1. Push  . The Calibration counter is shown, for example: Cal Count = 4.
2. Push  . The display shows Password: ????. The factory default password is 1234.
The top row of buttons corresponds to a number:
•  = 1
• = 2
•  = 3
•  = 4
As each number is entered, the cursor automatically moves to the next number.
3. To enter 1234, push , , , and  .
4. Push   (Start) to go to the first calibration point, C-01.
If you enter the correct password, C-01 shows on the display. If the password is incorrect, ???? shows on the 
display and you must enter the password correctly to go to the first calibration point, C-01. 
Change the Password
If you change the password and then lose it, see Restore the Default Password on page 14.
To change the password:
1. Push  . The Calibration counter is shown, for example: Cal Count = 4.
2. Push  . The display shows ????.
3. Enter the correct password digits. 
4. Push  (Change). The display shows ----.
5. Enter the new password digits. 
6. Push   (Save) to enter and store the new password.