8
Temperature Specifications
Temperature Resolution Accuracy
[1,2]
-200 °C to +1350 °C 0.1 °C
1 % + 10
-328 °F to +2462 °F 0.1 °F
1 % + 18
[1] Does not include error of the thermocouple probe.
[2] Accuracy specification assumes ambient temperature stable to ±1 °C. For ambient temperature changes of ±5 °C, rated accuracy
applies after 2 hours.
Capacitance and Diode Test Specifications
Function Range Resolution Accuracy
1 nF
[1]
0.001 nF 1 % + 5
10 nF
[1]
0.01 nF 1 % + 5
100 nF
[1]
0.1 nF 1 % + 5
1 μF 0.001 μF
1 % + 5
10 μF 0.01 μF
1 % + 5
100 μF 0.1 μF
1 % + 5
1000 μF 1 μF
1 % + 5
10 mF 0.01 mF 1 % + 5
Capacitance
100 mF 0.1 mF 2 % + 20
Diode Test 3.1 V 0.0001 V 1 % + 20
[1] With a film capacitor or better, using relative mode (REL Δ) to zero residual.
Frequency Counter Specifications
Function Range Resolution Accuracy
99.999 Hz 0.001 Hz 0.02 % + 5
999.99 Hz 0.01 Hz 0.005 % + 5
9.9999 kHz 0.0001 Hz 0.005 % + 5
99.999 kHz 0.001 Hz 0.005 % + 5
[1]
Frequency
(0.5 Hz to 999.99 kHz, pulse
width >0.5 μs)
999.99 kHz 0.01 Hz 0.005 % + 5
Duty Cycle
[2][3]
1.00 % to 99.00 % 0.01 % 0.2 % per kHz + 0.1 %
0.1000 ms 0.0001 ms 0.002 ms + 3 counts
1.000 ms 0.001 ms 0.002 ms + 3 counts
10.00 ms 0.01 ms 0.002 ms + 3 counts
Pulse Width
[2][3]
1999.9 ms 0.1 ms 0.002 ms + 3 counts
[1] For 64.000 kHz to 67.000 kHz, accuracy = ±5 Hz, at -20 to +55 °C with R.H. 0 % to 90 % (0 °C to 37 °C), 0 % to 65 % (37 °C to 45
°C), 0 % to 45 % (45 °C to 55 °C).
[2] For rise times <1 μs. Signals centered around trigger levels.
[3] 0.5 to 200 kHz, pulse width >2 μs. Pulse width range is determined by the frequency of the signal.

Table of Contents

Other manuals for Fluke 287

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 287 and is the answer not in the manual?

Fluke 287 Specifications

General IconGeneral
Data holdYes
Product typeDigital multimeter
Product colorGrey, Yellow
Frequency range0 - 999.99 kHz
DC current range0 - 10 A
DC voltage range0 - 1000 V
Resistance range0 - 500000000 Ω
Capacitance range0 - 100000 nF
Resolution (AC voltage)0.001 mV
Basic accuracy (frequency)0.005%
Basic accuracy (DC current)0.05%
Basic accuracy (DC voltage)0.025
Temperature measurement range-200 - 1350 °C
Battery typeAA
Power sourceBattery
Battery technologyAlkaline
Number of batteries supported6
Storage temperature (T-T)-40 - 60 °C
Operating temperature (T-T)-20 - 55 °C
Operating relative humidity (H-H)0 - 90 %
Display digits4.5 digits
Weight and Dimensions IconWeight and Dimensions
Depth60 mm
Width102 mm
Height222 mm
Weight870.9 g