287/289
Calibration Manual
26
Table 7. Calibration Adjustment Steps for Firmware Version 1.00 (cont.)
Step
Rotary Switch
Position
Source Value
Settling Time
(seconds)
Calibrating Time
[1]
(seconds)
Connect the 5520A in a 4-Wire configuration and select COMP 2-Wire
30 mV DC 500.0 mV, 60.0 Hz 5 3
Connect the 5520A in a 4-Wire configuration and select COMP 2-Wire
31 Ω 0 Ω, zcomp wire2 4 20
32 Ω 500 Ω, zcomp wire2 5 2
33 Ω 5 kΩ, zcomp wire2 5 2
34 Ω 50 kΩ, zcomp wire2 5 2
Disconnect 5520A 4-Wire configuration and turn COMP off
35 Ω 500 kΩ 5 5
36 Ω 5.0 MΩ 5 5
37 Ω 50.0 MΩ 5 5
38 Ω 500.0 MΩ 5 2
39 Diode Test 0 Ω, zcomp wire2 2 2
40 Diode Test 3.1 V, 0 Hz 2 2
41 μA DC 500.0 μA, 0 Hz 3 4
42 μA DC 5000.0 μA, 0 Hz 3 4
43 μA AC 500.0 μA, 60 Hz 10 4
44 μA AC 5000.0 μA, 60 Hz 10 3
45 mA DC 50.0 mA, 0 Hz 5 4
46 mA AC 400.0 mA, 0 Hz 3 4
47 mA AC 50.0 mA, 60 Hz 10 3
48 mA AC 400.0 mA, 60 Hz 10 4
49 ADC 5.0 A, 0 Hz 8 4
50 ADC 10.0 A, 0 Hz 3 2
51 AAC 5.0 A, 60 Hz 10 4
52 AAC 10.0 A, 60 Hz 9 4